Simplera™ CGM support
for use with the InPen™ smart insulin pen

Select the topic that fits your support needs:
Choosing your insertion site
The Simplera™ CGM is only approved to be worn in the back of the upper arm.

Do not use the Simplera™ CGM on the abdomen or other body sites including the buttocks, due to unknown or different performance that could result in hypoglycemia or hyperglycemia.
The sensor should be at least 1 inch away from any manual insulin injection site (syringe, pen).
For best sensor glucose performance, avoid these sites:
- Where clothing may rub or be too tight
- Where your body naturally bends a lot as this may cause the sensor to pull out
- Where there’s a lot of movement or rubbing (such as the inside of your arm)
- That are scarred, have hardened tissue, or stretch marks
Note: Assistance from another person may be needed for sensor insertion into the back of the upper arm. If assistance is not needed, a mirror may be helpful for self-insertion.
Dos and Don'ts for sensor insertion
Do
- Push on top of the inserter (either with the palm of the hand or fingers) to activate the inserter
- Insert with a smooth, firm, continuous motion
Don't
- Wrap fingers around the inserter (with no force applied to the top)
- Go slowly, halfway, pause/stop right before activation

How to insert the Simplera™ CGM
The following steps can help you get the best performance and longest wear.
- Choose an insertion site on the back of your upper arm that’s free of scar tissue.
- Inspect the inserter to ensure the tamper band is intact, then twist the cap to open.
-
Prep the skin using an alcohol wipe and let your arm completely dry before inserting.
-
Press the inserter firmly yet swiftly against the insertion site until you hear a click.
-
Use your finger to smooth down the adhesive.

Starting a new Simplera™ sensor
When your sensor expires, your app will notify you and a prompt will appear.
Tap Simplera™ to begin
On the Select device screen, tap Simplera™.
Confirm your sensor by tapping I have this sensor.
The next screen will tell you to not throw away the inserter after you insert the sensor. You will need the serial number and code that are printed on the top of the inserter. Tap Next on this screen.
Next, insert your sensor, then tap I inserted it.
Your app will search for the sensor. On the Select the sensor screen, select the serial number that matches the one you just inserted.
Look for the 6-digit code on the inserter. Enter the code on the Enter code screen, then tap Confirm.
You should now see a Sensor Paired! message.
It will take up to 2 hours for your sensor to warm up before you start to see sensor glucose (SG) values on the home screen. For the first 12 hours after inserting a new sensor, use a blood glucose meter reading to make treatment decisions.
Tips to manage bleeding
Bleeding may occur at the insertion site and is not unusual. Here are 3 tips to help you manage bleeding:
- If bleeding persists, apply steady pressure with a sterile gauze pad or clean cloth on top of the sensor for up to 3 minutes.
- If bleeding continues after 3 minutes and is significantly visible, or there is excessive pain or discomfort, then remove the sensor and continue to apply steady pressure until bleeding stops. Discard sensor and insert a new Simplera™ CGM in a different spot!
- Check the site for redness, bleeding, irritation, pain, tenderness, or inflammation. If you notice any of these, contact a healthcare professional.