Troubleshooting the MiniMed Mobile app


Device: MiniMed 780G insulin pump

First steps:

The MiniMed Mobile app is a secondary display system only. Always keep in mind that the information on the app may look different from user to user depending on which compatible mobile device is paired.

If you are unable to resolve an issue after you follow the steps in the tables below, force the app to close and then restart the app. If you still encounter an issue, restart your compatible mobile device and restart the app.

If you are using an Android operating system, you may need to try one or more of the following solutions:

  • Set the correct date and time on the compatible mobile device.
  • Enable Google services.
  • Clear cache and data from Google services.
  • Clear cache and data from the Google Play store.
  • Update Google services.
  • Install any pending operating system updates.

MiniMed Mobile app error messages

The following is a table of app error messages. These messages are generated by the app itself, not the paired pump or other system devices.
Note: If the app is locked or unresponsive, close and restart the app.

Error message Instructions
No Internet Connection
An Internet connection is needed to open the app.
Please retry when an Internet connection is available.
This message occurs during the initial setup of the app. The mobile device must be connected to the Internet to open the app for the first time.
Connect the compatible mobile device to the Internet using a wireless connection or enable cellular data for the app in the cellular data settings on your compatible mobile device.
For information on how to manage cellular data settings on your compatible mobile device, refer to the user guide for that device. If your compatible mobile device is connected to the Internet, the CareLink system may be down. Before you contact 24-Hour Technical Support or the local Medtronic support representative, try restarting your compatible mobile device and opening the app again. If that does not work, contact 24-Hour Technical Support or the local Medtronic support representative.
App Reset Required
An error has occurred.
The app must reset to continue. You will have to pair your pump again.
Before you attempt to pair the app again, remove the compatible mobile device from the list of paired devices on the pump. You must also remove the pump from the list of paired devices on the compatible mobile device.
Tap OK to begin the setup process.
The startup sequence that you saw the first time you opened the app begins. Follow the instructions to set up the app.
Unsupported Android or iOS Device
Unfortunately the app does not work with your mobile device.
The app cannot be used on the mobile device it was downloaded to.
Unsupported Android or iOS Version
Unfortunately the app does not work with the current Android or iOS version on your mobile device.
The app cannot be used with the current operating system on the mobile device it was downloaded to.
New Android or iOS Version
You can still use the app!
However, the app has not been tested with the current Android or iOS version on your mobile device. Information may display differently than intended.
You can still use the app but information may not appear as intended.
Tap Continue to open the app.
Pump Communication Error
An error has occurred in communication with the pump. Try turning Bluetooth off and back on. Alternatively, try restarting your mobile device.
Tap Continue. First turn Bluetooth off and back on. You may need to restart your compatible mobile device.
App Cannot Be Used
In order for the MiniMed Mobile app to work properly, we do not allow the app to run on a device with modified operating system (rooted device).
Tap Close The App to close the app. Use a compatible mobile device that does not have a modified operating system.
App Cannot Be Used
In order for the MiniMed Mobile app to work properly, we do not allow the app to run on a device with modified operating system (jailbroken device).
Tap Close The App to close the app. Use a compatible mobile device that does not have a modified operating system.
Developer Options Enabled
In order for the MiniMed Mobile app to work properly, we do not allow the app to run on a device with Developer Options enabled.
To use the app, disable Developer Options.
Tap Settings to access your compatible mobile device operating system settings. Disable Developer Options in the Settings on your compatible mobile device.
Oops, something went wrong
Unable to connect to our servers, check your Internet connection. If Internet connection is working, our servers may be temporarily not responding.
Please try again later.
Tap Retry to try to connect again. Tap Cancel to try again later.
App cannot be used
To continue setup, app notifications are required to be enabled. Please allow notifications in settings.
Tap Settings and allow notifications from the app.
Secure Screen Lock Needed
For your security, MiniMed Mobile app requires you to lock your mobile device with a PIN, pattern, biometric or password.
Add a secure authentication method for unlocking your personal mobile device, such as a PIN, pattern, biometric ID, or password.


MiniMed Mobile app status messages

The following is a table of app status messages.
Note: If the app is locked or unresponsive, close and restart the app.

App message Instructions
Bluetooth off Turn on the Bluetooth™* wireless feature on your compatible mobile device to see your current sensor information.
Pump pairing lost There were no data exchanges with the pump for several minutes. Pair the compatible mobile device with the pump again. Before you attempt to pair the app again, remove the compatible mobile device from the list of paired devices on the pump. You must also remove the pump from the list of paired devices on the compatible mobile device.
Updating...
This may take a few minutes
The app is receiving and processing new sensor data. This may take a few minutes and then the app shows the data received.
Reconnecting to pump... Move the pump closer to the compatible mobile device. Keep the pump and the compatible mobile device within 20 feet or 6 meters of each other without obstacles.


CareLink manual upload failed messages

Error message Instructions
CareLink Upload Failed.
CareLink might be temporarily not responding. Please try again later. If this problem persists, call Helpline.
Try again later. If this problem persists, contact 24-Hour Technical Support or the local Medtronic support representative.
Pump Out of Range
Pump is out of range. Move pump closer to connected mobile device and try again.
Move your pump closer to the connected mobile device and try again.
Server Upload Failed
The server might be temporarily unavailable. Please try again later. If this problem persists, call Helpline.
Try again later. If this problem persists, contact 24-Hour Technical Support or the local Medtronic support representative.
Pump Communication Error
An error has occurred in communication with the pump. Try turning Bluetooth off and back on. Alternatively, try restarting your mobile device.
First turn Bluetooth off and back on. You may need to restart your compatible mobile device.
Pump not Paired
The upload could not be performed since the pump is not paired with the mobile device.
Pair the pump with your compatible mobile device.
Check Internet Connection
Check Internet connection and please try again later.
Check your internet connection and try again later.

The MiniMed Mobile app drains my battery
If the pump and the MiniMed Mobile app lose communication frequently, you may experience battery drain on both the pump and the compatible mobile device when you use the app.

To minimize battery drain, always make sure the compatible mobile device and the pump are within 20 feet or 6 meters of each other without obstacles.

No SG value in the status area
If the sensor feature is turned off on your pump, the MiniMed Mobile app displays the current active insulin value where the SG value is usually displayed in the Status area. If the sensor feature is turned on in the pump settings, the app will display a status message if it is unable to display the current SG value.

The pump does not pair with the MiniMed Mobile app
If the pump does not pair with the MiniMed Mobile app on your compatible mobile device, any of the following actions may help resolve the issue.

  • Confirm that you have removed any previously paired pumps from Bluetooth™* settings menu on your compatible mobile device.
  • Confirm that you have removed any previously paired compatible mobile devices from the Paired Devices menu on the pump.
  • Turn the Bluetooth™* setting off and then on again on your compatible mobile device.
  • Close and then reopen the MiniMed Mobile app.
  • Restart your compatible mobile device.


The MiniMed Mobile app does not sync with my CareLink account

Issue Possible solution
You have no Internet connection. Return to an area where you can connect your compatible mobile device to the Internet.
The CareLink server is down. Try again later.
The app is not running on the compatible mobile device. The MiniMed Mobile app is forced to close and cannot run in the background. The compatible mobile device settings must let the app run in the background for the app to continue to sync to CareLink and receive data from the pump. Since this setting cannot be managed in the app, refer to your compatible mobile device instructions to find out how to manage this setting.
The compatible mobile device is applying a battery saving feature to the app, and this is preventing the app from syncing with your CareLink account. Turn off the battery saving feature on your compatible mobile device for the app. Since this setting cannot be managed in the app, refer to your compatible mobile device instructions to find out how to manage this setting.
Sync to CareLink is disabled. From the Menu, tap Sync to CareLink. Tap the toggle at the top of the screen so that the switch turns green. You have enabled the automatic Sync to CareLink feature. You can also perform a manual upload to your CareLink account by using the Upload Now feature.
The app is not receiving data from the pump. Check the app for any status messages or icons that tell you why communication between the pump and the app was lost.
There are no reports available in my CareLink account even though the app is synced with the CareLink account, and other data appears in the CareLink account. The Sync to CareLink feature has been enabled for less than 24 hours, either because the app has just been set up, or because the connection between the app and your CareLink account has not been functional for other reasons. The data history that CareLink needs to generate reports has not been uploaded to your CareLink account. Use the Upload Now feature to send data from the app to your CareLink account.