ALERT: Due to the inclement weather across much of the country, Medtronic deliveries are currently experiencing delays and there will be longer than normal hold times on the phone. Visit www.Diabetes.shop if you need to place an order, get a sensor or belt clip replacement, or check the status of your supply order. X

Returning to SmartGuard™ Auto Mode

MiniMed 770G System Support   >   Returning to SmartGuard™ Auto Mode

Returning to SmartGuard™ Auto Mode


Device: MiniMed™ 770G Insulin Pump

If you have automatically transitioned to Manual Mode, you can return to Auto Mode if all readiness conditions are satisfied and you enter a blood glucose (BG).

You can return to Auto Mode if the following conditions are satisfied:

  • Auto Mode is enabled on your pump.
  • Your sensor is providing good sensor glucose (SG) values.
  • A bolus is not in progress.
  • A temp basal rate is not in progress.
  • 48-hour warm-up is complete.
  • Auto Mode is not in a five-hour warm-up period.
  • You have entered a new BG reading.

If any of these conditions are not met, Auto Mode cannot restart.