ALERT: Due to the inclement weather across much of the country, Medtronic deliveries are currently experiencing delays and there will be longer than normal hold times on the phone. Visit www.Diabetes.shop if you need to place an order, get a sensor or belt clip replacement, or check the status of your supply order. X

Exiting SmartGuard™ Auto Mode

MiniMed 770G System Support   >   Exiting SmartGuard™ Auto Mode

Exiting SmartGuard™ Auto Mode


Device: MiniMed™ 770G Insulin Pump

The pump will exit SmartGuard™ Auto Mode:

  • Auto Mode has been in Safe Basal for 90 minutes.
  • A High SG alert has occurred.
  • You have not cleared any suspend event messages within four hours.
  • You manually turned off the Sensor feature or disconnected the transmitter.

Some alarms cause the pump to exit SmartGuard™ Auto Mode and also turn the Auto Mode feature off. Auto Mode is turned off if an alarm initiates a pump reset. If this occurs, you will no longer see the SmartGuard™ Auto Mode shield on your Home screen. You must turn the Auto Mode feature on again and go through a five-hour warm-up period.

You can turn off Auto Mode at any time.